Measuring range:
Resolution 0,1 / 0,01 pH - 0,01 μS
Automatic and manual temperature compensation
DHS digital sensors management.
Supplied as standard: Electrode holder stand, colored buffer solutions, power supply, and instructions for use. Without pH electrode. With 2301T cell, NT 55 temperature probe, S7-BNC cable, and accessories.
Need a customised solution?
Optional accessories are available for purchase separately. If your order needs a customised configuration, our technical specialists will follow up with full details before your order is finalised. Contact us via the enquiry form or sales@lasec.com.
Installation, training and special delivery
Some products require installation, training, or special delivery (such as tail-lift trucks, extra labour, or hazardous material packaging). If your order needs this, we'll contact you with the additional cost before dispatch, so your final invoice may differ from the listed price.
Ordering from outside South Africa
Online ordering is currently limited to South African customers, but our full range is still available. Log an enquiry or email international@lasec.com and our international sales team will assist you through our traditional sales channels.
Description: Professional bench meters with innovative backlit high-definition colored LCD display, with the possibility of brightness adjustment. Possibility to select the parameters to show on the display. The step-by-step guide in setup and calibration allows simple and immediate use of the device. Automatic pH calibration with USA and NIST buffers, up to 3 points and 2 points defined by the user. Automatic ORP calibration on 1 point. Automatic conductivity calibration up to 4 points (84; 147; 1413 μS and 12,88; 111,8 mS) and 1 point custom. Automatic resolution, reference temperature from 15 to 30 °C, and selectable cell constant (C= 0,1; 1; 10). Indication of the calibrated points, with icons. Measurement stability indication and the possibility to select 3 stability levels. All operations are constantly monitored and shown to the operator by the colored LED above the display and self-diagnosis messages.
| REGULATIONS | |
| Regulations | CE |
| PERFORMANCE | |
| Accuracy (Cond) | ± 2 % f.s. |
| Accuracy (pH) | ± 0.02 |
| Accuracy (Temperature) | ± 0.5 °C |
| Cond measuring range (max.; min.) | 200 mS; 0.00 μS |
| Cond resolution | 0.01 / 0.1 / 1 μS/cm (automatic scale) |
| ORP measuring range (max.; min.) | + 1000; - 1000 |
| ORP resolution | 1 mV |
| pH measuring range (max.; min.) | + 14.00; 0 |
| pH resolution | 0.01 |
| Reference temperature (max.; min.) | + 30 °C; + 15 °C |
| TDS Factor (max.; min.) | 1.00; 0.40 |
| TDS measuring range (max.; min.) | 200 g/l; 0.1 mg/l |
| Temperature coefficient (max.; min.) | 10.00 % / °C; 0.00 % / °C |
| Temperature compensation | 0.00...80.0 °C |
| Temperature measuring range (max.; min.) | + 100.0 °C; 0.0 °C |
| Temperature resolution | 0.1 °C |
| Working temperature (max.; min.) | + 45 °C; 0 °C |
| FEATURES | |
| Calibration report | Yes |
| DHS compatible | Yes |
| Display | Backlight coloured high-definition LCD |
| GLP calibration | Yes |
| Inputs | BNC - RCA |
| IP protection | IP 54 |
| Points of calibration Cond | 1...4 |
| Points of calibration ORP | 1 |
| Points of calibration pH | 1...3 auto / 2 custom values |
| Printer connection | Yes |
| Selectable Cell constant | 0.1 / 1 / 10 |
| Stability criteria | 3 levels |
| Standard approved Cond | 84 μS - 147 μS - 1413 μS - 12.88 mS - 111.8 mS - 1 user value |
| Standard approved pH | USA - NIST - 2 custom values |
| POWER SUPPLY | |
| Power supply | 100 - 240 V / 50 - 60 Hz |
| DIMENSION AND WEIGHT | |
| Dimension (LxWxH) | 160 x 140 x 45 mm |
| Weight | 380 g |
| Brand | Giorgio Bormac s.r.l |
|---|