Steriliser 108L, SF110

Steriliser 108L, SF110

R67,847.82
  • Volume: 108 L

  • Performance Class: SingleDISPLAY
  • Air Circulation: Forced air circulation (Fan)

  • Setting temperature range: +20 to +250 °C
  • Function SetpointWAIT: the process time does not start until the set temperature is reached

  • Door: fully insulated stainless steel door with 2-point locking (compression door lock)

  • Stainless steel interior: w(A) x h(B) x d(C): 560 x 480 x 400 mm (d less 39 mm for fan); easy-to-clean interior, made of stainless steel, reinforced by deep drawn ribbing with integrated and protected large-area heating on four sides

  • Textured stainless steel casing: w(D) x h(E) x d(F): 745 x 864 x 584 mm (d +2x56mm door handle); rear zinc-plated steel

  • Voltage - Electrical load: 230 V, 50/60 Hz; approx. 2500 W; 115 V, 50/60 Hz; approx. 1800 W

  • If you should require a customised special solution, please contact our technical specialists using our enquiry form.
  • Special delivery requirements, such as the need for tail-lift trucks and additional labour and installation for large equipment and hazardous material (Hazmat) packaging, will attract additional freight charges (quoted separately).
Availability:On Request
SKU
IMEMSF110
BrandMemmert

Steriliser S

Sterilisation done without requiring disinfection of laboratory instruments and equipment. A hot air sterilisation device protects and saves lives in medicine.

Standard bearer in hot air sterilisation

What is hot air sterilisation? Steriliser S makes use of dry sterilisation to ensure laboratory equipment and devices are sterilized. With this method, hot air, devoid of water vapor or moisture, makes sure the device is sterile without needing risk prone disinfection methods. Each Memmert hot air steriliser S is classified as a class IIb medical device. Memmert is bringing medical devices of class IIb into circulation according to MDD 93/42/EEC until May 26, 2024, according to the transitional provisions of the (EU) 2017/745 Article 120 (2).

  • Hot air sterilisation temperature ranges up to +250 °C
  • 8 model sizes (32 to 749 liters volume) to heat sterilize as needed
  • 2 model variants: SingleDISPLAY andTwinDISPLAY
  • Natural convection or forced air circulation N/F

Advantages of using hot air sterilisation in Steriliser S

  • Intuitive and user friendly operating menu
  • SetpointWAIT function ensures that the program is only continued when the setpoint hot air sterilisation temperature is reached
  • Optional process-controlled, electromagnetic door lock (models with TwinDISPLAY)
  • User-ID function prevents manipulation in hot air sterilisers like Steriliser S
  • Almost exclusive use of high-quality, corrosion-resistant and easily cleanable stainless steel for the working chamber and housing
  • Heating concept specifically adapted for a precise and homogenous temperature control
  • A wide range of options for reading the data logging using interfaces, integrated data logger and Software AtmoCONTROL

Importance of hot air sterilisation

  • Disinfection of appliances and instruments is not sufficient in medicine and research. Only hot air sterilisation provides sufficient protection.
  • Precisely maintaining the heat sterilisation time and terminating even the most resistant microorganisms. These are guaranteed at all times, irrespective of the hot air sterilisation load and chamber volume.
  • For the special application of depyrogenisation, the hot air Steriliser S is absolutely ideal as an example of hot air sterilisation.

The indispensable safety feature for this hot air steriliser is SetpointWAIT. This means that the programme for sterilisation only starts when the set temperature has been reached. This feature can also be used with freely positionable Pt100 temperature sensors. Here, the sterilisation time only begins when the set temperature has been reached at all measurement points, and reliable sterilisation is guaranteed at all times.

Temperature
Setting temperature range+20 to +250 °C
Working temperature rangeat least 10 above ambient temperature to +250°C
Setting accuracy temperatureup to 99.9 °C: 0.1 / from 100 °C: 0.5
Temperature sensor1 Pt100 sensor DIN class A in 4-wire-circuit
Control technology
ControlCOCKPITSingleDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with high-definition TFT-colour display
Language settingGerman, English, Spanish, French, Polish, Czech, Hungarian
adjustable parameterstemperature (Celsius or Fahrenheit), fan speed, air flap position, programme time, time zones, summertime/wintertime
TimerDigital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function SetpointWAITthe process time does not start until the set temperature is reached
Calibrationthree freely selectable temperature values
Communication
Documentationprogramme stored in case of power failure
ProgrammingAtmoCONTROL software for reading out, managing and organising the data logger via Ethernet interface (temporary trial version can be downloaded). USB stick with AtmoCONTROL software available as accessory (on demand).
Safety
Temperature controladjustable electronic overtemperature monitor and mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature
Autodiagnostic systemfor fault analysis
Ventilation
Fanforced air circulation by quiet air turbine, adjustable in 10 % steps
Fresh airAdmixture of pre-heated fresh air by electronically adjustable air flap
Ventvent connection with restrictor flap
Standard equipment
Works calibration certificateCalibration at +160°C
Doorfully insulated stainless steel door with 2-point locking (compression door lock)
Internals2 stainless steel grid(s), electropolished
Stainless steel interior
Dimensionsw(A) x h(B) x d(C): 560 x 480 x 400 mm (d less 39 mm for fan)
Interioreasy-to-clean interior,made of stainless steel, reinforced by deep drawn ribbing with integrated and protected large-area heating on four sides
Volume108 L
Max. number of internals5
Max. loading of chamber175 kg
Max. loading per internal20 kg
Textured stainless steel casing
Dimensionsw(D) x h(E) x d(F): 745 x 864 x 584 mm (d +2x56mm door handle)
Housingrear zinc-plated steel
Electrical data
Voltage
Electrical load
230 V, 50/60 Hz
approx. 2500 W
Voltage
Electrical load
115 V, 50/60 Hz
approx. 1800 W
Ambient conditions
Set UpThe distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm.
Ambient temperature+5 °C to +40 °C
Humidity rhmax. 80 %, non-condensing
Altitude of installationmax. 2,000 m above sea level
Overvoltage categoryII
Pollution degree2
More Information
Brand Memmert
Write Your Own Review
You're reviewing:Steriliser 108L, SF110
Your Rating