Model: Incubator IN260F0
Volume: 256 l
Dimensions: w(A) x h(B) x d(C): 640 x 800 x 500 mm
Performance Class: SingleDISPLAY
Air Circulation: Natural air circulation (Convection)
Includes: Entry port, 23 mm clear diameter, for introducing connections at the side, can be closed by flap, position left centre/centre (Order no. F0)
If you should require a customised special solution, please contact our technical specialists using our enquiry form or at sales@lasec.com.
Special delivery requirements, such as the need for tail-lift trucks and additional labour and installation for large equipment and hazardous material (Hazmat) packaging, will attract additional freight charges (quoted separately).
Incubator I
The incubator I is perfect for the world of research, medicine, pharmaceutics, and food analytics, as well as food chemistry. The heating of the IN260 Standard Delivery Incubator is optimally tuned for natural convection, and valuable chamber loads for research, pharmaceutics, medicine, and food chemistry are warmed up very carefully.
Key Features:
Intuitive Operation: Touch, turn & go! Easy and intuitive operation with the unique Memmert ControlCOCKPIT and software AtmoCONTROL.
Stainless Steel: Beautiful and functional. Memmert stainless steel is scratch-resistant, hygienic, robust and ever-lasting.
Unlimited Communication: Memmert appliances are equipped with state-of-the-art interfaces for programming and communication.
Safety Of Users and Chamber Loads: electronic temperature monitoring system, mechanical temperature limiter as well as an integrated auto-diagnostic system.
Highly sensitive, extremely accurate Incubator
In the laboratory, the Memmert incubator I is ideal for all applications with temperatures up to +80°C – specifically for incubating living cultures at +37 °C. Thanks to the finely-tuned control technology, critical temperature overshoots are completely ruled out, valuable loads are therefore warmed up particularly carefully in this highly precise microbiological incubator.
Memmert incubator I – a gentle force in the laboratory
To guarantee optimum air circulation in the incubator chamber even when it is fully loaded, both model variants SingleDISPLAY and TwinDISPLAY are available with an adjustable air fan. Since Memmert’s unique all-round surface heating is optimally geared to both natural convection and forced air circulation, the fan can be switched off entirely for sensitive samples and for samples which may not be exposed to air turbulence.
A Memmert incubator offers unique advantages:
Temperature | |
---|---|
Setting temperature range | +20 to +80 °C |
Working temperature range | at least 5 above ambient temperature to +80 °C |
Setting accuracy temperature | 0.1 °C |
Temperature sensor | 1 Pt100 sensor DIN class A in 4-wire-circuit |
Control technology | |
Language setting | German, English, Spanish, French, Polish, Czech, Hungarian |
ControlCOCKPIT | SingleDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with high-definition TFT-colour display |
Timer | Digital backwards counter with target time setting, adjustable from 1 minute to 99 days |
Function SetpointWAIT | the process time does not start until the set temperature is reached |
Calibration | three freely selectable temperature values |
adjustable parameters | temperature (Celsius or Fahrenheit), air flap position, programme time, time zones, summertime/wintertime |
Ventilation | |
Convection | natural convection |
Fresh air | Admixture of pre-heated fresh air by electronically adjustable air flap |
Vent | vent connection with restrictor flap |
Communication | |
Documentation | programme stored in case of power failure |
Programming | AtmoCONTROL software for reading out, managing and organising the data logger via Ethernet interface (temporary trial version can be downloaded). USB stick with AtmoCONTROL software available as accessory (on demand). |
Safety | |
Temperature control | adjustable electronic overtemperature monitor and mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature |
Autodiagnostic system | for fault analysis |
Standard equipment | |
Works calibration certificate | incl. works calibration certificate for +37°C |
Door | fully insulated stainless steel door with 2-point locking (compression door lock) |
Door | inner glass door |
Internals | 2 stainless steel grid(s), electropolished |
Stainless steel interior | |
Interior | easy-to-clean interior,made of stainless steel, reinforced by deep drawn ribbing with integrated and protected large-area heating on four sides |
Volume | 256 L |
Dimensions | w(A) x h(B) x d(C): 640 x 800 x 500 mm |
Max. number of internals | 9 |
Max. loading of chamber | 300 kg |
Max. loading per internal | 20 kg |
Textured stainless steel casing | |
Dimensions | w(D) x h(E) x d(F): 824 x 1183 x 684 mm (d +56mm door handle) |
Housing | rear zinc-plated steel |
Electrical data | |
Voltage Electrical load | 230 V, 50/60 Hz approx. 1700 W |
Voltage Electrical load | 115 V, 50/60 Hz approx. 900 W |
Ambient conditions | |
Set Up | The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm. |
Altitude of installation | max. 2,000 m above sea level |
Ambient temperature | +5 °C to +40 °C |
Humidity rh | max. 80 %, non-condensing |
Overvoltage category | II |
Pollution degree | 2 |
Brand | Memmert |
---|