Climate and Humidity Chamber, 56L, Model HCP50

R 181,705.61
  • Chamber type: Climate and humidity chamber with active humidity control
  • Working temperature range: Ambient +7 °C up to +90 °C
  • Setting temperature range: +18 °C to +90 °C
  • Setting accuracy (temperature): 0.1 °C 
  • Humidity range: 20–95 % rh (active microprocessor control)
  • Setting accuracy (humidity): 0.5 % rh 
  • Water supply: Demineralised water only; conductivity 5–10 µS/cm; pH 5–7; external tank with self-priming pump (tank and hose included)
  • Internal volume: 56 L
  • Internal dimensions (W × H × D): 400 × 425 × 330 mm (usable depth less 35 mm for fan)
  • External dimensions (W × H × D): 559 × 795 × 521 mm (depth +56 mm for door handle)
  • Net weight: 55 kg
  • Electrical supply: 230 V or 115 V, 50/60 Hz; approx. 1 520 W
  • Interfaces: Ethernet LAN, USB
  • Certifications: CE, VDEGS, cMET, EAC
  • Calibration document: Works calibration certificate included (60 °C / 75 % rh, chamber centre)

Need a customised solution?

Optional accessories are available for purchase separately. If your order needs a customised configuration, our technical specialists will follow up with full details before your order is finalised. Contact us via the enquiry form or sales@lasec.com.

Installation, training and special delivery

Some products require installation, training, or special delivery (such as tail-lift trucks, extra labour, or hazardous material packaging). If your order needs this, we'll contact you with the additional cost before dispatch, so your final invoice may differ from the listed price.

Ordering from outside South Africa

Online ordering is currently limited to South African customers, but our full range is still available. Log an enquiry or email international@lasec.com and our international sales team will assist you through our traditional sales channels.

DISCLAIMER: All product images are for illustrative purposes only. Actual products may vary slightly in colours, controls, and visible features.
Availability:On Request
SKU
IMEMHCP50
BrandMemmert

Product Overview

Memmert Climate and Humidity Chamber, HCP Series

Active humidity control for environmental simulation and accelerated service life testing

The Memmert HCP series comprises four climate chambers designed for applications that require precise, simultaneous control of temperature and relative humidity. Active humidity regulation across the full range of 20–95 % rh, combined with all-round surface heating on six sides including a heated inner glass door, produces a stable interior environment with minimised condensation risk and high homogeneity of both temperature and humidity.

Available in four sizes from 56 to 241 litres, the HCP series is suited to environmental simulation, accelerated service life testing, and 85/85 tests in accordance with IEC 60068-2-67 and IEC 60068-2-78. An aluminium thermal conductive layer supports uniform temperature distribution and serves as a heat accumulator during temporary power interruptions. Double doors are standard across all models, with a wide-area heated inner glass door that limits contamination risk and temperature fluctuation during access while providing a clear view of the chamber load.

All models are operated via the Memmert ControlCOCKPIT with TwinDISPLAY and the AtmoCONTROL software, which supports programming and documentation of temperature and humidity profiles. The SetpointWAIT function ensures that process runtimes begin only once the set temperature is reached, and three-fold overtemperature protection is standard. A rear panel feed-through port is included on all models. Battery-buffered ControlCOCKPIT and MobileALERT remote alarm notification (e-mail and SMS) are available as options.

  • Max. Temperature: +90 °C
  • Humidity range (rh): 20–95 %
  • Available volumes: 56–241 L
  • 4 models: HCP50 to HCP240

Features

  • Active humidity control: Maintains 20–95 % rh with short recovery times after door opening and no routine maintenance requirements.
  • All-round surface heating: Six-sided heating including a heated inner glass door minimises condensation and reduces vapourisation inside the chamber.
  • Double door — standard: Outer door and wide-area heated inner glass door reduce contamination risk while allowing visual inspection of the chamber load without opening.
  • Aluminium thermal conductive layer: Supports uniform temperature distribution and acts as a heat accumulator during temporary power interruptions.
  • ControlCOCKPIT and AtmoCONTROL: Touch-and-turn TwinDISPLAY interface with software for programming and documenting temperature/humidity profiles.
  • SetpointWAIT function: Process runtime begins only once the setpoint temperature is reached, ensuring test conditions are met before timing starts.
  • Three-fold overtemperature protection: Simultaneous display of all parameters and triple safety system protects samples throughout the run.
  • Rear panel feed-through: Standard port on all models for cabling and sensor connections without compromising the chamber seal.

Applications

  • Environmental testing and simulation
  • Accelerated service life testing
  • 85/85 tests per IEC 60068-2-67 and IEC 60068-2-78
  • Climate-controlled sample conditioning

Available Models

Model

Volume (L)

Internal W × H × D (mm)

HCP50

56

400 × 425 × 330

HCP105

107

560 × 480 × 400

HCP150

156

560 × 700 × 400

HCP240

241

600 × 810 × 500

Temperature
Working-temperature range+7 above ambient temperature up to +90 °C
Setting temperature range+18 to +90 °C
Setting accuracy temperature0.1 °C
Temperature sensor2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
Humidity
Humidity ControlActive microprocessor control for humidifying and dehumidifying (20 – 95 % rh), incl. digital indication and auto-diagnostic system ensures even more rapid reaching of set humidity and very short recovery times. Humidity supply with water (only for demineralised water with a conductivity of 5 to 10 μS/cm and a pH value between 5 and 7; from an external tank) by a self-priming pump; integral bacteria block by generating hot steam, dehumidifying via sterile filter
Setting range active humidity control20 to 95 % rh and rh-Off
Setting accuracy humidity0.5 % rh
Control technology
ControlCOCKPITTwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
Language settingGerman, English, Spanish, French, Polish, Czech, Hungarian, Italian, Chinese
adjustable parameterstemperature (Celsius or Fahrenheit), relative humidity, programme time, time zones, summertime/wintertime
TimerDigital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function SetpointWAITthe process time does not start until the set temperature is reached
Calibrationthree freely selectable values each, temperature and humidity
Communication
InterfaceEthernet LAN, USB
Documentationprogramme stored in case of power failure
ProgrammingAtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port
Safety
AutoSAFETYadditionally integrated over- and undertemperature monitor "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature
Autodiagnostic systemintegral fault diagnostics for temperature and humidity control
Alarmvisual and acoustic
Heating concept
6 sideslarge-area multi-function heating system on four sides with additional door and back heating to avoid condensation
Standard equipment
Works calibration certificateWorks calibration certificate (measuring point chamber centre) at 60 °C with 75 % rh
DoorAdditional heated inner glass door
Doorfully insulated stainless steel door with 2-point locking (compression door lock)
Entry portEntry port (silicone), 40 mm clear diameter, for introducing connections, moisture tight, can be closed by silicone stopper, at the back, centre left
Internals1 perforated stainless steel shelf/shelves
Standard accessoriesWater tank including connection hose
Stainless steel interior
Dimensionsw(A) x h(B) x d(C): 400 x 425 x 330 mm (d less 35 mm for fan)
InteriorW.-St. 1.4301 (ASTM 304)
Volume56 l
Max. number of internals
Max. loading of chamber75 kg
Max. loading per internal15 kg
Textured stainless steel casing
Dimensionsw(D) x h(E) x d(F): 559 x 795 x 521 mm (d +56mm door handle)
Electrical data
VoltageElectrical load230 V, 50/60 Hz
approx. 1520 W
Voltage Electrical load115 V, 50/60 Hz
approx. 1520 W
Ambient conditions
Ambient temperature10 °C to 30 °C
Set UpThe distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm.
Humidity rhmax. 70 %, non-condensing
Altitude of installationmax. 2,000 m above sea level
Overvoltage categoryII
Pollution degree2
Packing/shipping data
Transport informationThe appliances must be transported upright
Customs tariff number8419 8998
Country of originFederal Republic of Germany
WEEE-Reg.-No.DE 66812464
Dimensions approx incl. cartonw x h x d: 730 x 950 x 640 mm
Net weightapprox. 55 kg
Gross weight cartonapprox. 74 kg