Note: UV-translucent plates (P1PLA051C-000096), Computer to operate the device and to analyse results (ILSCCOMPI5), Installation and operational training quoted separately
Online ordering is currently limited to South African customers, but our full range is still available. If you are ordering from outside South Africa, please log an enquiry or email internationallabsales@lasec.com, and our international sales team will gladly assist you through our traditional sales channels.
The SPECTROstar Nano is equipped with BMG LABTECH’s ultra-fast UV/vis spectrometer, which allows users to measure full-spectrum absorbance (220 – 1000 nm) in less than 1 second per well. It is the ideal instrument to read all absorbance assays in a microplate or cuvette – with many common assay protocols predefined and available at the click of the mouse.
FEATURES:
UV/Vis spectrometer-based absorbance up to 4 OD. Full absorbance spectrum (220 to 1000 nm) for each well at a resolution of 1 nm.
Cuvette port for standard and low-volume measurements
Accurate and precise incubation up to 45C
Reads 6-, 12-, 24-, 48-, 96-, 384-, well plates
Orbital, double orbital, and linear shaking
Automated path length correction
Quick Start plate reading feature
Multi-User Reader Control and MARS Data Analysis Software package included (can be installed by an unlimited number of users at no extra cost)
FDA 21 CFR part 11 compliant software
OPTIONAL ACCESSORIES QUOTED SEPARATELY:
680-101 LVis Plate (Absorbance plate with filters allowing operational qualification, containing a horizontal cuvette port and equipped with 16 low volume positions for absorbance measurements down to 2uL sample volumes)
680-102 LVis Plate (Absorbance plate containing a horizontal cuvette port and equipped with 16 low volume positions for absorbance measurements down to 2 uL sample volumes)
601-650 Measurement up to 1536-well plate format
601-750 Gas vent connection for atmospheric sensitive samples
INSTALLATION OF THIS INSTRUMENT QUOTED SEPARATELY:
Installation + operational training will comprise unpacking the unit, installing the software, and performing a fluorescence PQ using fluorescein.
Must provide UV-translucent plates or purchase separately.
200ul pipette + tips + 1.5ml tubes + water are necessary for the procedure.