Online ordering is currently limited to South African customers, but our full range is still available. If you are ordering from outside South Africa, please log an enquiry or email internationallabsales@lasec.com, and our international sales team will gladly assist you through our traditional sales channels.
The GR Series Intelligent Autoclaves offer a comprehensive solution for sterilisation needs. This system features a sizable, heavy-gauge sterilisation chamber, satisfying a range of sterilisation requirements. Its microcomputer control system provides precise oversight throughout the sterilisation process, ensuring both accuracy and ease of maintenance. Notably, the autoclave includes a waste sterilisation mode tailored for lab wastes and adheres to the CE standard for drying. A straightforward control panel simplifies operation, and safety is enhanced with a unique mechanical interlocking system. The GR Series combines functionality with user-centered design to meet diverse sterilisation demands.
Heavy-Gauge Sterilisation Chamber
The sterilisation chamber has a diameter of 40cm, providing capacity suitable for various sterilisation tasks.
Microcomputer Control System
This system uses a microcomputer control mechanism to manage the sterilisation process. It also supports maintenance, inquiry, calibration, recording, and scalability features. The autoclave accommodates three tiers of administration for efficient use and management.
Waste Sterilisation Mode
The autoclave is equipped with protocols specifically for sterilising laboratory waste.
Drying
The drying function complies with the CE standard.
Cooling Lock OPEN Temperature
Users have the option to set different cooling lock OPEN temperatures for each sterilisation process. This is to account for the thermal inertia of items being sterilised, preventing premature opening of the lid, especially with liquids that may still be at high temperatures.
Simple Control Panel
The control panel is designed for clarity and ease of operation.

Preset Procedures and Memory System
Six preset procedures are available for various sterilization and warming tasks, including the treatment of solids and liquids and the handling of agars. The system also allows for the creation and storage of up to 60 custom cycles.
Safety
The autoclave features a mechanical interlocking system with an "eight-column evenly distributed" design. Additionally, electric double inner locks are in place, with the control system monitoring the temperature and pressure within the sterilisation chamber. The chamber can only be unlocked when safe temperature and pressure levels are detected.
| Model | Standard + cooling fan |
| Capacity (L) | 60 |
| Dimension(L*W*H,mm) | 660x644x980 |
| Chamber dimension(Dia*H,mm) | Ø400x505 |
| Rate power supply (w) | 2900 |
| Stainless steel basket(Dia*H,mm) | (Ø360×200)x2 |
| Power requirement | 220V,16A,50/60HZ |
| Temperature accuracy | 0.1℃ |
| Working environment | 5-45℃,relatively humidity 10%-85% |
| Chamber material | SUS 304 |
| Sterilisation temperature | 105-138℃, designed temp: 152℃ |
| Sterilisation time range | 1-6000 minutes |
| Melting temperature | 60-100℃ |
| Melting time range | 1-6000 minutes |
| Warming temperature | 45-60℃ |
| Warming time range | 1-9999 minutes |
| Cooling lock OPEN temperature | 40-80℃ for liquid cycle, 40-99℃ for other cycles |
| Sterilisation cycle | 1. Liquid 2. Liquid with warming 3. Solid 4. Wrapped instrument 5. Fabric 6. Rubber 7. Fast 8. Waste 9. Agar melting 1 O. Self defined 11. Drying(only for drying models |
| Sterilisation cycle number | Can save up to 20 cycles |
| Exhaust mode | Automatic internal discharging, with a steam collecting bottle inside. Exhaust speed can be selected by 6 levels |
| Lid opening way | Up and down |
| Steam collector and water tank | Embedded 2 steam collector |
| Preset range of timer | Can be preset and auto start up to 6 days in advance |
| Display range of pressure gauge | -0.1-0.5Mpa |
| Pressure | Max working pressure:0.27Mpa,design pressure:0.35Mpa |
| Controller | “Inspiration” fast speed microcomputer controller |
| Safety device | “eight column evenly distributed” interlocking device, electric double inner locks, lid closing checking, dry scorch protection, overpressure protection, safety valve, water level sensor, over temperature protection, temperature monitor, over current and short circuit protection, leakage protection, cooling lock, anti-scald chamber cover and bench, automatic troubleshooting |
| Brand | Zealway |
|---|